Programmable AC power source for Flux system

Cet article n'est plus disponible.

Conditions générales
Lead time 6-8 weeks, built to order 


  • Max Power: 500VA
  • Size: 4U
  • Weight: 30kg
  • Input Voltage (50/60Hz): 1Ø2W 110V ±10% 1Ø2W 220V ±10%


Model we use in our flux system with the integrating sphere   Programmable AC power source
Phase Model Max Power Output Current Size Weight Input Voltage (50/60Hz)
0~150V 0~300V
ASP-CF-0500EP1P 500VA 4.2A 2.1A 4U 30kg 1Ø2W 110V ±10% 1Ø2W 220V ±10%

series Programmable AC power source

General Specifications :      
Output frequency FIx : 50Hz / 60Hz / 400Hz ; Adjustable : 40.00~500.0Hz
Frequency resolution 0.01Hz (40.00-99.99Hz) , 0.1Hz(100.0-999.9Hz)
Frequency accuracy ±0.01 %
Frequency regulation Fix : ±1Count ; Adjustable : ±0.01Hz at 40.00-99.99Hz , ±0.1Hz at 100.0-999.9Hz
Frequency temperature effect 50 ppm / ℃
Frequency response 40Hz~500Hz ( ±1db )(at output voltage 80% Full scale no load)
External input level 40.00Hz ~500.0Hz  ( 3.0Vrms Full scale) (optional)
Output voltage Standard : 0~150V / 0~300V Optional : 0~150V / 0~300V / 0~600V (Special voltage on request)
Voltage temperature effect ±0.2% / ℃
Output T.H.D. ≦0.25% (at 120V-50/60Hz ) (0.1%/100Hz) ( at Resister Load ,PF=1,THD-R measurement ) (to 80kHz BW)
Load regulation ±0.2% (at 50/60Hz , Resistor Load , PF=1 ,0.2%/100Hz)
Line regulation ±0.2% (ACV± 10% ; THD≦2% ; Frequency ± 5%) (at 80% fullscal ; Load=0~50% ; PF=1)
Allow crest factor of load current 2.5(at full load) (Rated current)
Max. peak current 5 times of Rated Current (1000mS)
Power factor of load 0.5~1(lead or Lag)
Output voltage ramp time 0.1sec ~ 99.9sec
Line drop compensation ≦ 5V
Set accuracy Voltage ≦0.1% ± 2C (at 50/60Hz ; 0.05% / 100Hz) Frequency ≦0.05% ± 1C (at 50/60Hz ; 0.05% / 100Hz)
Transient Response 100 micro sec
Circuit Linear Amplifier Type
Efficiency >50% (at 80%fullscal voltage ; Rated current) (watt efficiency)
S/N Ratio More then 75dB(At Rated Output Voltage)
DC drift < ±1mV
Protections Function
  • Over voltage protection ,
  • Over current protection(120%) ,
  • Line input over/under voltage protection (±15%) ,
  • ±Sense cable loosen protection or compensation voltage> 5V ,
  • Over temperature protection ,
  • Output waveform CLIP alarm
Cooling system Forced air-cooling type
Operating Temperature/Humidity 0 ~ +45 ℃ / 10 ~ 90%RH
Insulation Resistance 10Mohm .500VDC(line input terminal to casing)
Acoustic Noise 70~80dB At 1 Meter Distance
 
Functions Resolution (4 digital) Accurate
Frequency Meter 0.01Hz / 0.1Hz Auto Range ±0.1% reading ± 2C ± 0.05%/100Hz
True RMS Voltage Meter 0.1V ±0.2% reading ± 0.3% Full scale ± 2C ± 0.1% / 100Hz
True RMS Current Meter 1mA / 10mA or 10mA / 100mA  Auto Range ±0.4% reading ± 0.3% Full scale ± 3C ± 0.1% / 100Hz
Watt Meter 0.1W / 1W  or  1W / 10W  or  10W / 100W   Auto Range ±0.2% reading ± 0.2% Full scale ± 5C (at 50/60Hz) ±0.3% reading ± 0.2% Full scale ± 5C ± 0.1%/100Hz
VA Meter 0.1VA / 1VA or  1VA / 10VA or  10VA / 100VA  Auto Range ±0.05% of VA range
PF Meter 0.001

--