LTV 1200 Invasive Ventilator
This product is no longer available.
Peak inspiratory Pressure 0-120 cmH2O
Disconnect/Sense
High and low O2 inlet pressure
Ventilation modes:Control, Assist/Control, SIMV/CPAP and NPPV Standards and regulatory compliance
This product is not approved by Health Canada
The ventilation tools you need throughout the continuum of care for patients 5 kg and above
The LTV® 1200 ventilator is an essential part of an emergency preparedness plan and can help provide a ventilation response to mass casualty incidents—from influenza outbreaks to biological warfare attacks to natural disasters.
Compact, deployable design
The small, lightweight LTV 1200 ventilator provides sophisticated critical care ventilation in austere environments. The ventilator supports both invasive and noninvasive applications. It combines portability, impact-resistant construction, easyto-read displays and comprehensive monitoring. A single solution for any patient weighing 5 kg or more, the LTV 1200 ventilator weighs only 14.5 lbs and is a pickup-and-go solution that can travel with patients from field care to transport to the intensive care unit (ICU). It can take advantage of any 12-volt power source from cigarette lighters to outlets—and its internal battery allows it to be transferred from one power source to the next without interruption. Intuitive controls allow first responders to deploy the system with minimal retraining.
Quick patient ventilation
Choose infant, pediatric or adult to quickly configure initial ventilation before optimizing settings for each patient’s needs. This feature aids simple and effective emergency set-up.
Repurpose ease
While the LTV 1200 ventilator can be stockpiled for use in emergencies, many customers use it in hospital settings. More versatile than transport ventilators and significantly smaller than ICU systems, the LTV 1200 ventilator has intuitive controls
and comprehensive monitoring capabilities that can play an essential role in day-today healthcare delivery and mass-casualty medical care.
Oxygen (O2) resource management
The O2 Conserve feature reduces O2 consumption when the time and place of care require it. This feature helps extend the sometimes scarce resource of O2 during transport or emergencies and may be turned off when O2 resources are again in full supply.
SprintPack Lithium-Ion Power System
The SprintPack Lithium-Ion Power System is equipped with advanced technology designed to empower its users with comfort, confidence and convenience. Lithium-ion technology provides the greatest degree of power density in the lightest weight package of any battery type. This provides LTV 1200 ventilator users longer battery power in a smaller and lighter package. The SprintPack Lithium-Ion Power System also provides a higher number of full recharging cycles that extend the battery life to provide greater long-term value.
Customizable system configurations
The LTV 1200 ventilator can be custom-configured with a wide range of accessories to meet the needs of demanding care environments.
Accessories | Part number | |||||
LTM graphics monitor package | 17650-001 | |||||
Floor stand | 10 611,0 $ | |||||
Short crossbar | 11 451,0 $ | |||||
Long crossbar | 11 452,0 $ | |||||
Circuit support arm | 11 501,0 $ | |||||
Humidifier pole | 11 453,0 $ | |||||
Utility basket | 11 450,0 $ | |||||
SprintPack Lithium-Ion Power System | 19222-001 | |||||
SprintPack bracket | 19096-001 | |||||
SprintPack transport pack | 19102-001 | |||||
Specifications
c | ||||||
Weight | 14.5 lb (6.5 kg) | |||||
Height | 3.25” (8.4 cm) | |||||
Width | 10.5” (27 cm) | |||||
Depth | 13.5” (38 cm) | |||||
Variable alarms | ||||||
Apnea interval | 10–60 sec | |||||
High pressure limit | 5–100 cmH2O | |||||
Low peak pressure | Off, 1–60 cmH2O | |||||
Low minute volume | Off, 0.1–99 L | |||||
High PEEP | Off, 3–20 cmH2O above set PEEP | |||||
Low PEEP | Off, -3– -20 cmH2O below set PEEP | |||||
High rate | Off, 5–80 BPM | |||||
Alarm volume | 60–85 dBA |
Controls | |
Power | On/Standby |
Ventilation modes | Control, Assist/Control, SIMV/CPAP and NPPV |
Breath types | VC, PC, PS and spontaneous |
Breath rate | 0–80 BPM |
Tidal volume | 50–2,000 mL |
Inspiratory time | 0.3–9.9 sec |
PC/PS/Spontaneous flow | 160 LPM |
Pressure control | 1–99 cmH2O |
Pressure support | 1–60 cmH2O |
Sensitivity | Off, 1–9 LPM |
O2% | 21–100% |
O2 flush | 1–3 min |
Low pressure O2 | On/Off |
PEEP/CPAP | 0–20 cmH2O |
Insp/Exp hold | 6 seconds maximum |
Manual breath | 1x current settings |
Control lock | Easy or hard unlock options |
Monitors and indicators | |
Peak inspiratory pressure | 0–120 cmH2O |
Mean airway pressure | 0–99 cmH2O |
PEEP | 0–99 cmH2O |
Breath rate | 0–250 BPM |
Airway pressure display | -10–108 cmH2O |
Exhaled tidal volume | 0–4,000 mL |
Exhaled minute volume | 0–99.9 L |
I:E ratio | 99:1–1:99 |
Calculated peak flow | 10–100 LPM |
AutoPEEP | 0–100 cmH2O |
Static compliance | 1–999 mL/cmH2O |
Patient effort | Green LED |
Fixed alarms | |
Disconnect/Sense | |
External power low and lost | |
High and low O2 inlet pressure | |
Internal battery low and empty | |
Ventilator inoperative |
Extended features | |
SBT | |
Ventilator presets (infant, pediatric and adult) | |
Variable rise time | |
Variable flow termination | |
Variable time termination | |
Pressure control flow termination | |
Leak compensation | |
O2 conserve | |
O2 cylinder duration calculation | |
O2 flush | |
Apnea interval | |
Pneumatic specifications | |
High pressure O2 source | 40–80 PSIG (2.8–5.5 bar) |
Low pressure O2 source | < 80 LPM, < 10 PSIG |
Power indicators | |
External power | Green and amber LEDs |
Battery charge status | Green, amber and red LEDs |
Battery level | Green, amber and red LEDs |
Power specifications | |
Ventilator input 11–15 VDC | |
AC adapter input 100–250 VAC | |
50–60 Hz | |
Environmental specifications | |
Operating temperature | 5–40 °C (40–104 °F) |
Storage temperature | -20–60 °C (-4–140 °F) |
Operating humidity | 15–95% relative |
Storage humidity | 10–95% relative |
Shock compliance | MIL-STD-810F |
Vibration compliance | MIL-STD-810F |
Standards and regulatory compliance | |
RTCA/DO-160F | |
ANSI/UL STD 60601-1 | |
IEC 60601-2-12 | |
cETLus | |
CAN/CSA C22.2 STD NO. 601.1 |