- Max Power: 500VA
- Size: 4U
- Weight: 30kg
- Input Voltage (50/60Hz): 1Ø2W 110V ±10% 1Ø2W 220V ±10%
Model we use in our flux system with the integrating sphere | Programmable AC power source | |||||||
Phase | Model | Max Power | Output Current | Size | Weight | Input Voltage (50/60Hz) | ||
0~150V | 0~300V | |||||||
1Ø | ASP-CF-0500EP1P | 500VA | 4.2A | 2.1A | 4U | 30kg | 1Ø2W 110V ±10% 1Ø2W 220V ±10% |
series Programmable AC power source |
General Specifications : |
Output frequency | FIx : 50Hz / 60Hz / 400Hz ; Adjustable : 40.00~500.0Hz |
Frequency resolution | 0.01Hz (40.00-99.99Hz) , 0.1Hz(100.0-999.9Hz) |
Frequency accuracy | ±0.01 % |
Frequency regulation | Fix : ±1Count ; Adjustable : ±0.01Hz at 40.00-99.99Hz , ±0.1Hz at 100.0-999.9Hz |
Frequency temperature effect | 50 ppm / ℃ |
Frequency response | 40Hz~500Hz ( ±1db )(at output voltage 80% Full scale no load) |
External input level | 40.00Hz ~500.0Hz ( 3.0Vrms Full scale) (optional) |
Output voltage | Standard : 0~150V / 0~300V Optional : 0~150V / 0~300V / 0~600V (Special voltage on request) |
Voltage temperature effect | ±0.2% / ℃ |
Output T.H.D. | ≦0.25% (at 120V-50/60Hz ) (0.1%/100Hz) ( at Resister Load ,PF=1,THD-R measurement ) (to 80kHz BW) |
Load regulation | ±0.2% (at 50/60Hz , Resistor Load , PF=1 ,0.2%/100Hz) |
Line regulation | ±0.2% (ACV± 10% ; THD≦2% ; Frequency ± 5%) (at 80% fullscal ; Load=0~50% ; PF=1) |
Allow crest factor of load current | 2.5(at full load) (Rated current) |
Max. peak current | 5 times of Rated Current (1000mS) |
Power factor of load | 0.5~1(lead or Lag) |
Output voltage ramp time | 0.1sec ~ 99.9sec |
Line drop compensation | ≦ 5V |
Set accuracy | Voltage ≦0.1% ± 2C (at 50/60Hz ; 0.05% / 100Hz) Frequency ≦0.05% ± 1C (at 50/60Hz ; 0.05% / 100Hz) |
Transient Response | 100 micro sec |
Circuit | Linear Amplifier Type |
Efficiency | >50% (at 80%fullscal voltage ; Rated current) (watt efficiency) |
S/N Ratio | More then 75dB(At Rated Output Voltage) |
DC drift | < ±1mV |
Protections Function |
|
Cooling system | Forced air-cooling type |
Operating Temperature/Humidity | 0 ~ +45 ℃ / 10 ~ 90%RH |
Insulation Resistance | 10Mohm .500VDC(line input terminal to casing) |
Acoustic Noise | 70~80dB At 1 Meter Distance |
Functions | Resolution (4 digital) | Accurate |
Frequency Meter | 0.01Hz / 0.1Hz Auto Range | ±0.1% reading ± 2C ± 0.05%/100Hz |
True RMS Voltage Meter | 0.1V | ±0.2% reading ± 0.3% Full scale ± 2C ± 0.1% / 100Hz |
True RMS Current Meter | 1mA / 10mA or 10mA / 100mA Auto Range | ±0.4% reading ± 0.3% Full scale ± 3C ± 0.1% / 100Hz |
Watt Meter | 0.1W / 1W or 1W / 10W or 10W / 100W Auto Range | ±0.2% reading ± 0.2% Full scale ± 5C (at 50/60Hz) ±0.3% reading ± 0.2% Full scale ± 5C ± 0.1%/100Hz |
VA Meter | 0.1VA / 1VA or 1VA / 10VA or 10VA / 100VA Auto Range | ±0.05% of VA range |
PF Meter | 0.001 |
--