Respiratory Ventilator T7 for Emergency & Transportation
Cet article n'est plus disponible.
High-end ventilator for emergency and transportation
Color touchscreen for simpler operation and navigation
O2 monitoring system integrated into the main unit
Easy to install, configure, adjust and extend modules in real situations
This product is not approved by Health Canada
The T7 Emergency and Transport Ventilator adopts the cutting-edge technology of the O2 monitoring system integrated into the main unit which makes it much efficient, more accurate and more time-saving than the traditional O2 sensor which requires the second time plug-in.
T7 Model Characteristics | ||
Function Items | Model Configuration | |
IPPV | ( or VCV, Intermittent Positive Pressure Ventilation) | Standard |
V-A/C | ( Volume— Assist / control ) | Standard |
V-S I M V | ( Volume Control — Synchronized Intermittent Mandatory Ventilation) | Standard |
P CV | ( Pressure Controlled Ventilation) | Standard |
P-A/C | ( Pressure—Assist / control ) | Standard |
P -S I M V | ( Pressure Control— Synchronized Intermittent Mandatory Ventilation) | Standard |
CPAP | ( Continuous Positive Airway Pressure) | Standard |
C PA P+PSV | ( CPAP + Pressure Support Ventilation) | Optional |
Ventilation mode | ||
BiPPV | ( Bi—Level Positive Airway Pressure Ventilation) | Optional |
APRV | ( Airway Pressure Release Ventilation) | Optional |
PRVC | (Pressure Regulated Volume Control) | Optional |
P RV C + PSV | (PRVC + Pressure Support Ventilation) | Optional |
B i P PV+PSV | (BiPPV + Pressure Support Ventilation) | Optional |
CPR | ( Cardiopulmonary Resuscitation) | Optional |
Other functions | ||
RSA | ( Respiratory Sinus Arrhythmia ) | Optional |
EtCO2 | ( End—Tidal CO2) | Optional |
Ventilator Control Parameters | ||
Control Parameters | Items | Model T7 |
Respiratory frequency | Adjusting range | 0,1~bpm-120bpm |
Adjusting step length | 1 bpm | |
Accuracy | ±1 Bpm or ±5% | |
Inspiration/expiration ratio | Adjusting range | 59:1-1:99 |
Adjusting step length | 0,1 | |
Accuracy | 4:1-1:4 range±10% | |
Tidal volume | Adjusting range | 20mL-2500mL |
Adjusting step length | 5mL | |
Accuracy | ±15m1 or ±15% | |
Oxygen concentration | Adjusting range | 40%,100% |
Adjusting step length | 10% | |
Accuracy | ±10%(v/v) | |
Suction pressure | Adjusting range | 3cmH20 —60cm H 20 |
Adjusting step length | 1cmH20 | |
Accuracy | ±2cmH20 or ±10% | |
Limited airway pressure | Adjusting range | 15cmH20-70cmH20 |
Adjusting step length | 1cmH20 | |
Accuracy | ±2cmH20 or±10% | |
Trigger pressure | Adjusting range | -20cmH20-20cmH20 |
Adjusting step length | 1 cmH2O | |
Accuracy | ±1cmH20 or±10% | |
Continuous positive airway pressure/ positive end-expiratory pressure CPAP/PEEP | Adjusting range | 0cmH20-30cmH20 |
Adjusting step length | 1 cmH2O | |
Accuracy | ±2cmH20 or±10% | |
Pressure support | Adjusting range | 0,3cmH20-35cmH20 |
Adjusting step length | 1 cmH2O | |
Accuracy | ±2cmH20 or±1 0% | |
Trigger flow | Adjusting range | 1L/min-15L/min |
Adjusting step length | 1 L/min | |
Accuracy | ±1 L/min or±15% | |
Pause time ratio | Adjusting range | 0%-80 % |
Adjusting step length | 5% | |
Accuracy | 10% | |
Pressure increase time | Adjusting range | 0.1s-2s |
Adjusting step length | 0.1s | |
Accuracy | ±0.2s or±20% | |
Expiration trigger sensitivity | Adjusting range | 5%-80 % |
Adjusting step length | 5% | |
Accuracy | 15% | |
Oxygen inhalation flow rate | Adjusting range | 2L/min-30L/min |
Adjusting step length | 1 L/min | |
Accuracy | ±1 L/min or±1 5% |
Other Performance Indicators | |
Items | Model T7 |
Max. ventilation flow rate | 150L/min(@450kPa) |
Sigh ventilation function | When setting the SIGH function, the tidal volume of sigh ventilation shall be 1.5-2.5 times of the tidal volume preset |
Inspiration/expiration resistance | When the flow is 30L/min and 60L/min respectively, inspiration/expiration resistance shall be no more than 6cmH2O |
Spontaneous respiration resistance | When the flow is 1 5L/min and3OL/min respectively, inspiration resistance shall be no more than6cmH2O |
Inspiration safety valve | When ventilator is powered off or out of gas, the patient may breath the fresh air through the safety valve |
Air circuit system safety pressure | Air circuit maximum safety pressure no more than 1 OOcmH2O |
CPR ventilation function | CPR operation/prompt function supported |
RSA ventilation function | RSA ventilation function supported |
Manual ventilation function | / |
System noise | 65dB(A) |
Multilingual Switch | Multilingual operation interface is switchable via menu. |
Battery Working hours | 10 hours |
Leakage Compensation | Yes |
Screen | 7" TET color screen Resolution:800x480 pixels With resistor type touch screen, 4 quick access touch buttons |
Operation Languages available | English, Spanish, French, German |
Power Specifications | Power adapter Input voltage: 100-240V— Input frequency: 50-60Hz Input Current: 0.7-1.5A |
Ventilator alarm parameters | |||
Alarm parameters | Items | T7 Model | |
Upper limit of minute ventilation | Setting Range | 1L/min-120L/min | |
Alarm level | High | ||
Lower limit of minute ventilation | Setting Range | 0L/min-119L/min | |
Alarm level | High | ||
Upper limit of airway pressure | Setting Range | 15cmH20-70cmH20 | |
Alarm level | High | ||
Lower limit of airway pressure | Setting Range | 0cnnH20-60cmH20 | |
Alarm level | High | ||
Respiratory frequency alarm | Upper limit | Setting Range | 10bpnn--125bpm |
Alarm level | Medium | ||
Lower limit | Setting Range | 0bpm-115bpm | |
Alarm level | Medium | ||
Suffocation alarm time | Setting Range | 5-60 s | |
Alarm level | High | ||
Upper limit of tidal volume | Setting Range | 50mL--3000m1 | |
Alarm level | Medium | ||
Lower limit of tidal volume | Setting Range | 0mL-2400m1 | |
Alarm level | Medium | ||
Upper limit of Et-0O2 | Setting Range | 1 mmHg-1 50mmHg | |
Alarm level | High | ||
Lower limit of Et-CO, | Setting Range | 0mmHg-149mmHg | |
Alarm level | High | ||
Gas supply pressure alarm | Gas supply pressure low | Alarm voice | Gas supply pressure below 270 kPa |
Alarm level | Medium | ||
Gas supply pressure high | Alarm voice | Gas supply pressure higher than 600kPa | |
Alarm level | High | ||
No gas supply pressure | Alarm voice | Gas supply pressure below 70 kPa | |
Alarm level | High | ||
System power supply alarm | Adapters falling off | Alarm voice | Adapter fails to be connected |
Alarm level | Medium | ||
Low battery | Alarm voice | Minimum work time between alarm and shutdown is 20min | |
Alarm level | Medium | ||
Insufficient battery capacity | Alarm voice | Battery level is running out | |
Alarm level | High | ||
Respiratory system integrity alarm | Alarm voice in different alarm levels | When respiratory circuit is off, ventilator shall within 3 respiratory cycles, trigger the "Respiratory System Fails to Connect Alarm", and give off the voice prompt of "Check Respiratory System & Settings". High level | |
Mute alarm | Period | <120s |
Other Ventilator Performance Indicators | ||
Monitoring parameters | Items | T7 Model |
Respiratory frequency | Monitoring range | 0bpm-120bpm |
Accuracy | ±1 Bpm or ±5% | |
Tidal volume | Monitoring range | OmL-3000mL |
Accuracy | ±15rnL or±15% | |
Minute ventilation | Monitoring range | OL/min-120L/min |
Accuracy | ±0.5L/min or±1 5% | |
Oxygen concentration | Monitoring range | 40%-100 % |
Accuracy | ±10%(v/v) | |
Airway pressure | Monitoring range | -100cmH20-20cmH20 |
Accuracy | ±2cmH20 or±1 0% | |
Et-CO2 | Monitoring range | 0mmHg-150mmHg |
Accuracy | (0-40 mmHg)±2mmHg; | |
(41-70 mmHg):±5%; (71-100 mmHg):±8%; | ||
(101-150 mmHg)±10% | ||
Wave display | Pressure waveform | Display pressure time waveform |
Flow waveform | Display flow time waveform | |
Volume waveform | Display volumetric time waveform | |
Ring diagram display | P-V Ring | Display pressure volume ring |
P-V Ring | Display flow volume ring diagram | |
P-V Ring | Display flow pressure ring diagram |
Accessories | |
2.603.00039 | Main Unit |
2.603.00042 | Standard accessory package |
5.001.00033 | Big hook for silicon mask |
5.001.00035 | Reusable siliicon mask for adult /4# |
5.001.00037 | Silicon headgear (for adult) |
2.609.00021 | Power adapter (EU Standard) |
1.124.00001 | Power cable |
2.601.00046 | Air source hose(Eight years life time) |
2.603.00028 | Battery |
2.603.00031 | Reusable respiratory circuit Adult (180 times) |
2.603.00036 | Flow sensor for adult |
10.001.00022 | Optional accessory package |
1.124.00005 | Power cable |
2.609.00020 | Power adapter (US Standard) |
2.603.00031 P | Reusable respiratory circuit Pediatric (180 times) |
2.603.00030 | Disposable respiratory hose |
5.000.00168 | Test lung (1L) |
5.001.00034 | Reusable siliicon mask for child /3# |
5.001.00046 | Reusable siliicon mask for child/2# |
2.603.00037 | Reusable Flow sensor for newborn (180 times) |
5.001.00049 | Tracheal intubation tube for adult/7# |
5.000.00674 | Tracheal intubation tube for adult/8# |
5.000.00675 | Tracheal intubation tube for adult/9# |
1.205.00080 | Disposable filter (HME) |
2.603.00040 | T7 Transport Platform (Includes the stretcher adaptor) |
2.601.00029 | Transport Carrying Case Package (Model ()(Includes the stretcher adaptor) |
1.205.00061 | Oxygen Regulator |
1.205.00072 | Preoxide Bridge |
2.613.00001 | Carrying case (bag only) |
5.001.00011 | 2.5L Oxygen Tank/Aluminium CGA870 |
2.606.00017 | Ambulance connection cable |
10.001.00063 | ETCO2 module |
5.000.00452 | Mainstream CO2 module |
5.000.00453 | Adult adapter (disposable) |
5.000.00454 | Newborn adapter (disposable) |
5.000.00457 | Cable fixing slot |